ANC300 Plug-In Modules
open loop control for attocube positioners
The ANC300 offers a completely modular design with up to seven slots for dedicated stepping
(ANM150), scanning (ANM200) or combined stepping and scanning (ANM300) modules.
ANM150
stepping module
Output Signals
stepping - voltage range 0 .. 150 V
stepping - frequency range 0 .. 10 kHz
stepping - maximum current 100 mA, 4.5 A peak (max. 1 ms)
resolution of signal generation 2.3 mV (16 bit)
output noise < 5 mVpp (20 MHz bandwidth)
maximum capacitive load 5 μF
zero voltage output in GND mode connected to chassis ground directly
output connectors BNC, 50 Ohm
ANM200 ANM300
scanning module stepping & scanning module
Output Signals
Output Signals
scanning - output voltage range 0 .. 150 V stepping - voltage range 0 .. 150 V
scanning - gain 15 stepping - frequency range 0 .. 10 kHz
scanning - maximum current 100 mA stepping - maximum current 100 mA, 4.5 A peak (max. 1 ms)
scanning - offset programmable offset scanning - output voltage range 0 .. 150 V
scanning - power bandwidth DC .. 500 Hz (1 µF load) scanning - gain 15
scanning - small signal bandwidth DC .. 100 kHz (1 µF load) via DC input, up
to 500 kHz via AC input scanning - maximum current 100 mA
scanning - offset programmable offset
scanning - slew rate 800 V/ms (1 µF load) scanning - power bandwidth DC .. 500 Hz (1 µF load)
resolution of signal generation 2.3 mV (16 bit) scanning - small signal bandwidth output noise < 5 mVpp (20 MHz bandwidth) DC .. 100 kHz (1 µF load) via DC input, up
to 500 kHz via AC input
output filter additional switchable output filter
(1.6, 16, 160, 1600 Hz) scanning - slew rate 800 V/ms (1 µF load)
resolution of signal generation 2.3 mV (16 bit)
maximum capacitive load 5 µF output noise < 5 mVpp (20 MHz bandwidth)
zero voltage output in GND mode connected to chassis ground directly maximum capacitive load 5 μF
output connectors BNC, 50 Ohm zero voltage output in GND mode connected to chassis ground directly
output connectors BNC, 50 Ohm
Input Signals
DC in voltage range -10..+10 V Input Signals AC in voltage range max. ±5 V AC DC in voltage range -10..+10 V
coupling DC, programmable offset AC in voltage range max ±5 V AC
input resistance 10 kOhm coupling DC, programmable offset
input connectors BNC, 50 Ohm input resistance 10 kOhm
input connectors BNC, 50 Ohm