attoPUBLICATIONS attoCFM | Page 2

attoCFM I low temperature confocal microscope, free-beam optics The attoCFM I has been developed to offer a maximum amount of flexibility for a convenient adaption to a large number of different quantum optics applications. This is realized by an external optics head, positioned outside the cryostat. Furthermore, the free-beam optical design allows for completely independent adjustment of the excitation and collection port. Therefore, applications such as Raman spectroscopy become accessible by appropriately filtering of the excitation and detection signals. The easy handling opens up new possibilities in quantitative surface char- acterization in the sub-micron range. The attoCFM I can optionally be equipped with an interferometric encoder for closed loop operation with 1 nm resolution, and an ultra large range scanner with 125 µm scan range at 4 K. 6 7 1 8 2 9 9 8 7 10 11 01 attoCFM I external optics head (see next page for details) 02 vacuum window 3 03 microscope insert 04 superconducting magnet (optional) 4 05 attoDRY1100 cryostat (optional) 06 attoFPSensor based closed loop sensors (optional) 5 07 low temperature compatible, high-NA apochromatic objective 08 quick exchange sample holder with 8 electrical contacts 09 ultra-large range xyz scanner 125 μm x 125 μm x 15 μm @ 4 K (optional) Schematic drawing of the low temperature attoCFM I and the attoDRY1100 cryostat (optional) 10 xyz coarse positioners 5 mm x 5 mm x 5 mm 11 ultra stable titanium housing